HARDWARE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SENSORS

 

 

 

 

 

 

SOFTWARE

 

 

 

 

 

 

 

UTILITIES &

DIMENSIONS

 

 

 

Reactor

 

 

 

 

 

 

 

 

Stopcock Actuators

 

 

Manifold Clamps

 

Syringe Drives

 

 

 

 

 

 

 

Vacuum Pump

 

 

Automation

 

 

 

 

 

 

 

 

Radioactivity

 

Pressure

 

Temperature

 

 

Graphical interface

 

 

 

Synthesis recipes

 

 

 

Compressed Air

 

Inert Gas

 

 

Case

 

Dimensions

 

 

 

• 10mL low metal contaminant Topas COC reactor or 10mL glass reactors

• Heating to 150°C when using a Topas COC reactor and 220°C when using glass reactors

• Contained compressed air cooling to ambient temperatures (all exhaust compressed air can be piped outside of the hotcell to eliminate contamination or pressurizing of the hotcell environment by the compressed air used to cool the reactor)

• Optional Vortex Tube compressed air cooling to approx. 8°C

 

• 12 x 3-position electric servo actuators

• Can rotate all the disposable manifold stopcocks to 3 positions: left, right and off

 

• Unique magnetically locking latches for disposable stopcock manifold clamping

 

• 2 x electrically actuated syringe drives

• Multi-syringe size capable: 1mL, 5mL, 10mL & 20mL sizes

• Force limiting drive control circuit stops driving the syringes plunger if the back-pressure is too high, and automatically resumes drive once the back pressure is reduced.  This specialized circuit is especially useful when eluting Ga-68 generators with high back pressures and delivers smooth & reliable generator elutions.

 

• Built-in chemically resistant vacuum pump, dual head

• Max vacuum -0.95bar

 

• Industrial PLC (Programmable Logic Controller) with wireless communications to the interface laptop or tablet PC

• PLC is housed in an external compact enclosure which is located outside of the hotcell to eliminate radiation damage to the electronics.  This ensures reliable operation even in high radiation fields.

• Synthesizer is connected to the PLC enclosure via 2 electrical cables

 

 

 

• 3 x collimated linear CsI(TI) crystal PIN diode radioactivity detectors

 

• 2 x pressure sensors for vacuum and inert gas pressure monitoring

 

• Thermocouple sensor for reactor temperatures

 

 

• Easy to use open platform operator interface, with sensor trends, historical data logging & analysis, synthesis reports, multi-level password protected user access, CFR 21 CFR Part 11 & GMP compliant

 

• Easily generated using unique Click-&-Save graphical recipe development technology and stored as Excel® step lists

 

 

• 6-8 bar (87-116 psi), 6mm O.D. push-in tube connection

 

• Helium, Nitrogen or Argon; 2-8 bar (29-116 psi), 1/8" O.D. tube compression connection

 

• Compact chemically resistance powder coated case

 

• 238 mm x 270/397 mm x 240 mm (WxHxD)

MULTISYN TECHNICAL SPECIFICATIONS

........................................................................................................................................................