HPLC+ TECHNICAL SPECIFICATIONS
HARDWARE
SENSORS
SOFTWARE
UTILITIES &
DIMENSIONS
Process Valves
HPLC Valve
HPLC Pump
HPLC UV Detector
Automation
Radioactivity
Pressure
HPLC Injection Fluid Detector
Graphical interface
Inert Gas
Case
Dimensions
• 3 chemically inert PEEK solenoid process valves
• Kalrez® seals, 2.5 bar, 1.4 mm orifice, 20 mL dead volume
• 6 port electric injector, 0.40mm bore, N60 S.S. stator, valcon H rotor
• 5mL or 10mL HPLC loop supplied
• Knauer HPLC pump with 10 or 50 mL/min pump head
• Optional quaternary L.P.G. & quad channel degasser
• Knauer UV/VIS detector with remote fibre optic flow cell
• Variable wavelength (190-750 nm)
• External Industrial PLC with Ethernet communications to interface computer
• 1 x tungsten collimated linear CsI(TI) crystal PIN diode radioactivity detector
• Analogue gauge and built in pressure regulator for inert gas pressure delivery
• 1 x software calibrated optical fluid detector with on-the-fly detection threshold adjustments
• Easy to use open platform operator interface, with sensor trends, historical data logging & analysis, synthesis reports, multi-level password protected user access, CFR 21 CFR Part 11 & GMP compliant
• Helium, Nitrogen or Argon; 2-8 bar (29-116 psi), 1/8" push-in tube connection
• Compact Acetal (POM)
• Easy installation (only one electrical cable connection to external PLC controller)
• 32.5 cm x 8.4 cm x 25.5 mm (WxHxD)
About iPHASE technologies
........................................................................................................................................................
iPHASE technologies is a world class designer and manufacturer of automated instrumentation for P.E.T. radiochemistry. By using the latest automation technologies coupled together with intelligent design and class leading software interfaces, our instrumentation is built by users for users.